 ESCO Horizontal Laminar Flow Cabinet
Home : Our Principals : ESCO : Biotechnology Equipment ESCO : ESCO Horizontal Laminar Flow Cabinet

ESCO Airstream« Horizontal Laminar Flow Cabinets

Esco Airstream« Horizontal Laminar Flow Cabinet offers proven protection for your samples and processes. With over thousands of units in use worldwide, this model offers a sensible balance of quality, performance features and cost-effectiveness. Like all Esco cabinets this model features many key innovations for which Esco is recognized for : mini-pleat separatorless ULPA filter technology, the best product protection in the world, external rotor motors, superior filter mechanical construction.

The intelligent blower system automatically compensates to maintain airflow as the filter loads (additional manual adjustment can be carried out to further prolong filter life). This unique features eliminates the need for constant speed control adjustments, while ensuring optimum performance and product protection.

MAIN FILTRATION AGENTS

  • Mini-pleat separatorless ULPA filter technology reduces energy consumption and delivers increased laminar airflow uniformity for better product and cross contamination protection. ULPA filters are double scan tested, at the time of manufacturing, and after installation.
  • ISO Class 3 air cleanliness within work zone as per ISO 14644.1 (equivalent to Class 1 as per US Federal Standard 209E, 100 times "cleaner" than the usual Class 100 classification on cabinets offered by the competition). All materials used in the product are cleanroom compatible.
  • High-quality polyester pre-filter and main ULPA fillter with a typical efficiency of 99.9997% at MPPS and 99.9998% at both 0.3 and 0.12 microns provide the best product protection in the world; typical ULPA filter lifespan is more than 3 years depending on the operating conditions.
  • Integral filter metal guard on the ULPA filter prevents accidental damage the filter media ; seamless . lter gasket is permanently molded on the . lter frame and will not deteriorate over time; aerosol (DOP/PAO) challenge test port included.

CONTROL FEATURES

  • Built-in solid state variable speed controller(s) (infinitely adjustable from zero to the maximum setting) with built-in RFI and noise filters is superior to conventional "step" controllers.
  • Standard cabinet control system consists of separate switches / indicator lights for blower, lights and UV light (UV light switch is interlocked with the light and blower switch for enhanced safety) plus a pressure gauge for airflow monitoring.

CONSTRUCTION FEATURES

  • Industrial-grade main body constructed of electro-galvanised steel; with an abrasion-resistant white oven-baked powder-coated finish.
  • Durable and easy to clean stainless steel work surface will never rust, chip, or generate particles; raised edge at back of work zone prevents spills from damaging the filter.
  • Permanently lubricated direct drive centrifugal blower(s); energy efficient external rotor motor type design reduces operating costs.
  • Extremely low noise and vibration levels (less than 61dBA at working position) due to proprietary construction and mounting technology.
  • Built-in warm white, electronically ballasted lighting offers excellent illumination throughout the work zone in order to reduce operator fatigue and is comfortable to the eyes. Light tubes are mounted out of the air stream for better airflow uniformity.
  • UV-resistant tempered glass sides increase operator comfort and provide a high level of protection against harmful UV rays when a UV lamp is installed; tempered glass is also stronger and more durable compared to plastic materials.
  • Designed to meet the safety requirements of IEC 61010-1 / EN 61010-1 / UL 61010A-1 / CSA C22.2 No. 1010.1-92. Components are UL listed / recognised.
  • Cabinet is shipped fully-assembled in wood crate; simply plug in the unit to a power source for operation - no local installation is required.
  • Extended warranty period of 3 years excluding consumable parts and accessories.

Technical Data

General Specifications

AHC-2AX

AHC-3AX

AHC-4AX

AHC-5AX

AHC-6AX

External Dimensions (Width x Depth x Height)

730 x 750 x 1105 mm

28.7" x 29.5" x 43.5"

1035 x 750 x 1105 mm

40.7" x 29.5" x 43.5"

1340 x 750 x 1105 mm

52.7" x 29.5" x 43.5"

1645 x 750 x 1105 mm

64.8" x 29.5" x 43.5"

1950 x 750 x 1175 mm

76.8" x 29.5" x 46.3"

Internal Work Zone (Width x Depth x Height)

570 x 600 x 570 mm

22.4" x 23.6" x 22.4"

875 x 600 x 570 mm

34.4" x 23.6" x 22.4"

1180 x 600 x 570 mm

46.6" x 23.6" x 22.4"

1485 x 600 x 570 mm

58.5" x 23.6" x 22.4"

1790 x 600 x 570 mm

70.5" x 23.6" x 22.4"

Air Volume (At Initial Velocity)

526 cmh (+/-59 cmh) 309 cfm (+/-35 cfm)

808 cmh (+/-90 cmh) 475 cfm (+/-53 cfm)

1090 cmh (+/-120 cmh) 641 cfm (+/-71 cfm)

1371 cmh (+/-153 cmh) 807 cfm (+/-90 cfm)

1653 cmh (+/-184 cmh) 972 cfm (+/-108 cfm)

Laminar Airflow Velocity

Average of 0.45 m/s or 90 fpm measured 150mm / 6" from filter face for 45 air changes / minute (uniformity is +/-20%)

Standards Compliance

  • Individually performance tested and certified at factory under controlled conditions forá :
  • General requirements: IEST-RP-CC002.2 and AS1386.5
  • Air cleanliness: ISO 14644.1 Class 3, IEST-G-CC1001, IEST-G-CC1002 and other equivalent air cleanliness requirements
  • Filter performance: IEST-RP-CC034.1, IEST-RP-CC007.1, IEST-RP-CC001.3 and EN1822
  • Electrical safety: IEC 61010-1 / EN 61010-1 / UL 61010A-1 / CSA C22.2 No. 1010.1-92

Air Cleanliness Within Working Area

ISO 14644.1 Class 3, US Federal Standard 209E Class 1 / M1.5, AS 1386 Class 1.5, JIS B9920 Class 3, BS5295 Class C, Class M10,000 as per KS 27030.1 and other equivalent cleanliness classifications of the VDI 2083 and AFNOR X44101

Main Filter Type

ULPA filter with integral metal guards and filter frame gaskets; fully compliant with EN1822 and IEST-RP-CC001.3 requirements

Main Filter Efficiency Ratings

Minimum: 99.9991% at 0.3Ám / 99.9985% at 0.12Ám / 99.9982% at MPPS

Typical : 99.9998% at 0.3Ám / 99.9998% at 0.12Ám / 99.9997% at MPPS

Pre-Filter

Disposable and non-washable polyester fibers with 85% arrestance / EU3 rated

Noise Level

Typically <61 dBA at initial blower speed setting measured as per IEST-RP-CC002.2

(figure based on 4 feet cabinet, subject to acoustic properties of test environment)

Light Intensity

>1600 lux / >149 foot candles, measured at work surface level (zero background) as per IEST-RP-CC002.2

Main Body Construction

1.5mmt / 0.06" / 16 gauge electro-galvanised steel with white oven-baked epoxy powder-coated fnish

Side Window Construction

Colourless and transparent UV-absorbing 5 mm / 0.2" tempered glass

Work Surface Construction

1.2mmt / 0.05" / 18 gauge stainless steel grade 304

Maximum Power Consumption /Current

220-240VAC / 50Hz 1Ph 584W / 2.54A 600W / 2.61A 614W / 2.67A 634W / 2.76A 1631W / 7.09A

110-130VAC / 60Hz 1Ph

828W / 6.90A

844W / 7.03A

858W / 7.15A

878W / 7.32A

1666W / 13.88A

Power Supply Options

Choose from the following options and specify option number when ordering (e.g. AHC-4A2 for a 110-130VAC 60HZ)

1. 220-240VAC 50HZ, 1 phase 2. 110-130VAC 60HZ, 1 phase 3. 220-240VAC 60HZ, 1 phase 4. 110-130VAC 50HZ, 1 phase 5. 100-110VAC 50HZ / 60HZ

Net Weight (Approximate)

92 kgs / 203 lbs

112 kgs / 247 lbs

133 kgs / 293lbs

161 kgs / 355 lbs

208 kgs / 443 lbs

Max Shipping Weight

185 kgs / 408 lbs

200 kgs / 440 lbs

223 kgs / 492 lbs

251 kgs / 553 lbs

297 kgs / 655 lbs

Max Shipping Dimensions (W x D x H)

900 x 950 x 1630 mm

35.4" x 37.4" x 64.2"

1200 x 950 x 1630 mm

47.2" x 37.4" x 64.2"

1500 x 950 x 1630 mm

59.0" x 37.4" x 64.2"

1800 x 950 x 1630 mm

70.9" x 37.4" x 64.2"

2100 x 950 x 1730 mm

82.7" x 37.4" x 64.2"

Max Shipping Volume

1.39 cbm / 49.1 cbf

1.86 cbm / 48.4 cbf

1.74 cbm / 65.7 cbf

2.79 cbm / 98.5 cbf

3.45 cbm / 121.8 cbf

NOTE: AHC-where A in the second part (XAX) of the code denotes the standard workzone height of 570mm / 22.4". The following alternative workzone heights are available upon request: B: 722 mm / 28.4" C: 875 mm / 34.4" ( e.g. AHC-or AHC-2CX). Specify code when ordering.

 Printable Version (View: 3783 | Print: 779)

Previous:
ESCO Labculture« Horizontal Laminar Flow Cabinets
Airstream« Horizontal Laminar Flow Cabinet (Stainless Steel Version)

ESCO Animal Handling Cabinet
ESCO Class II Safety Cabinet
ESCO Class III Biohazard Safety Cabinet
ESCO Custom Product
ESCO Cytotoxic Cabinet
ESCO Ergonomic Accessories
ESCO Formalin Vaporizer
ESCO Horizontal Laminar Flow Cabinet
ESCO IVF Cabinet
ESCO PCR Cabinet
ESCO Vertical Laminar Flow Cabinet
ESCO Weighing Cabinet
Options & Accessories
 
PT. PROLABMAS MURNI SWADAYA
Sentra Niaga Puri Indah Blok T3-10
Jakarta 11610 - Indonesia
Telp: +62 (21) 581 8000, 581 9000
Fax: +62 (21) 581 6999
 
Home  |  About Us  |  Products  |  Download  |  Support  |  Contact Us
 
   Back to top