ESCO Airstream® Class II Type B2 Biosafety Cabinets
Class II Type B2 biohazard safety cabinets provide product, operator and
environmental protection and are suitable for general microbiological work with
agents assigned to biological safety levels I, II or III. In a Class II Type B2
cabinet, all inflow and downflow air is exhausted after HEPA/ULPA filtration to
the external environment without recirculation within the cabinet.
These cabinets are supplied with an integral exhaust collar that allows the
cabinet exhaust to be ducted to the external environment via a dedicated exhaust
system.
As they incorporate a non-recirculating total exhaust design, Class II Type
B2 cabinets are suitable for work with toxic chemicals when these are present in
trace amounts.
MAIN FILTRATION AGENTS
- ISO Class 3 air cleanliness within workzone
as per ISO 14644.1 (equivalent to Class 1 as per the US Federal Standard 209E, 100
times cleaner than the usual Class 100 classification on cabinets
offered by the competition).
- Minipleat separatorless H13 HEPA / U15 ULPA filter
technology, operating at the typical efficiency of 99.999%
at MPPS, 0.3 and 0.12 microns provides better operator, product and
environmental protection than conventional HEPA filters.
CABINET CONTROL SYSTEM
- Esco Sentinel™ Microprocessor Cabinet Control System allows
the user to easily access all cabinet functions. The control system monitors and
displays cabinet airflow constantly on the control LCD screen, prompting the
user with audible and visual alarms in case of any unsafe condition.
- True airflow velocity sensing technology, with temperature compensation for
improved sensor accuracy.
- Audible and visual alarms for low airflow, unsafe sash positions.
- Automatic warm-up cycle is enabled upon turning on the cabinet, ensuring all
contaminants are purged from the cabinet workzone before the operator can use
the cabinet. All menu functions are inactive during this period. Automatic
post-purge cycle can be configured at shutdown, ensuring all residue
contaminants are purged out of the workzone before deactivation.
- Fail-safe system ensures that in case of exhaust failure, the cabinet’s
main fan automatically shuts down to ensure safety to the user and the
environment.
- An admin PIN can be set to restrict unauthorized access to all menu
functions.
ERGONOMIC FEATURES
- Attractive stainless steel internal side walls are
non-glaring and easy to clean. All joints are expertly sealed for optimum safety
and cleanability of the interior.
- Additional UV-interlock ensures UV lamp feature is deactivated when the sash
is not fully closed.
- Built-in 5000k fluorescent lighting offers
excellent illumination throughout the workzone. Electronically-ballasted
lighting system is instant-start, non-flickering and
energy-efficient.
- Sloped front design and frameless front sliding sash allow
for maximum visibility into the workzone.
- Cabinet armrest is raised above the workzone
to ensure that the operator’s arms do not block the front inflow perforations.
CABINET BODY CONSTRUCTION
- All components designed for maximum chemical resistance for long service life
and durability. All cabinet components are cleanroom compatible.
- Industrial-grade main body constructed of electrogalvanised steel: with an
abrasion-resistant white oven-baked powder-coated finish. Attractive
single-piece stainless steel work surface is easy to remove.
- Inherently safe design maintains containment
for protection even with work tray removed, thus ensuring safety for the
operator during cleaning.
- Permanently lubricated direct drive centrifugal blower(s); energy
efficient external rotor motor type design reduces operating costs;
extremely low noise and vibration levels (less than 64dBA at working position)
due to proprietary construction and mounting technology.
- Built-in solid state variable speed controller(s) (infinitely adjustable from
zero to the maximum setting) with built-in RFI and noise filters is superior to
conventional "step" controllers.
- Superior electrical safety for the operator; all electrical components are UL listed or UL recognized; all cabinets
are factory tested for electrical safety after production at manufacturing site.
- Designed to meet the general safety requirements of the IEC 61010-1 / EN
61010-1 / UL 61010A-1 / CSA C22.2 No. 1010.1-92.
- Cabinet is shipped fully-assembled; simply plug the unit into a power source
for operation - no local installation is required; unit is supplied with a 3-pin
plug; 10 international plug types are available (specify order code when
ordering).
Technical Data
|
General Specifications |
AB2-4SX* |
AB2-6SX* |
|
External Dimensions (L x W x H) |
1340 x 710 x 1460 mm / 52.8" x 28.0" x 57.5" |
1950 x 710 x 1460 mm / 76.8" x 28.0" x 57.5" |
|
Internal Work Zone (L x W x H) |
1270 x 550 x 670 mm / 50.0" x 21.7" x 26.4" |
1880 x 550 x 670 mm / 74.0" x 21.7" x 26.4" |
|
Standards Compliance |
EN 12469 :2000 (exceeds requirements for Class II microbiological
safety cabinets)
Air cleanliness : ISO 14644.1 Class 3, IEST-G-CC1001, IEST-G-CC1002 and
other equivalent requirements
Filter performance : IEST-RP-CC034.1, IEST-RP-CC007.1, IEST-RP-CC001.3
and EN1822
Electrical safety : IEC 61010-1 / EN 61010-1 / UL 61010A-1 / CSA C22.2
No. 1010.1-92 |
|
Average Airflow Velocities |
Inflow |
Initial setpoint: 0.45 m/s or 90 fpm (audible /
visual alarm will activate at 0.4m/s or 80fpm) |
|
Downflow |
Initial setpoint: 0.31 m/s or 61 fpm (uniformity is +/-20%) |
|
Airflow Volumes (At Initial Airflow Velocity Setpoints)
|
Inflow |
356 cmh / 210 cfm |
527 cmh / 310 cfm |
| Downflow |
738 cmh / 434 cfm |
1091 cmh / 642 cfm |
|
Total Exhaust |
1094 cmh / 644 cfm |
1618 cmh / 952 cfm |
|
Pressure Drop Requirements for Exhaust Ducting |
440 Pa / 1.77" WG |
365 Pa / 1.47" WG |
|
Cleanliness Within Working Area |
ISO 14644.1 Class 3, US Federal Standard 209E Class 1 / M1.5, AS 1386
Class 1.5, JIS B9920 Class 3, BS5295 Class C, Class M10,000 as per KS
27030.1 and equivalent classes of VDI 2083 and AFNOR X44101 |
|
Downflow Filter Type |
ULPA filter with integral metal guards and filter frame gaskets ; Typical efficiency 99.999% at 0.3µm, 0.12µm and MPPS; fully compliant
with EN 1822 and IEST-RP-CC001.3 requirements
|
|
Exhaust Filter Type |
HEPA filter with integral metal guards and filter frame gaskets ; Typical efficiency 99.99% at 0.3µm; fully compliant with EN 1822 and
IEST-RP-CC001.3 requirements
|
|
Power Supply Options |
*Choose from the following options and specify option number when
ordering, (e.g. AB2-4S1 for 220-240VAC 50HZ)
1. 220-240VAC 50HZ, 1 phase 2. 110-130VAC 60HZ, 1 phase 3. 220-240VAC